CJEU SnitemSoftware as a medical device
CJEU Snitem in brief
Software that uses patient data to detect, for example, contraindications or excessive doses is a medical device for that function, even without acting on the body. What counts is the manufacturer's medical purpose. Decided under Directive 93/42/EEC; under Art. 2(12) MDR, promotional materials also define the purpose.
CJEU Snitem in marketing approval
On 7 December 2017, on a reference from the French Conseil d’État, the CJEU ruled on when software is a medical device within the meaning of Directive 93/42/EEC. The case concerned drug prescription assistance software that France made subject to an additional national certification (paras 12 to 18).
Software is a medical device where it meets two cumulative conditions: a medical objective and an action within the meaning of the definition (para 22). Use in a medical context is not enough. The intended purpose defined by the manufacturer must be specifically medical (para 24). Software that cross-references patient-specific data with the drugs the doctor is contemplating prescribing and automatically detects, in particular, contraindications, drug interactions and excessive doses pursues such an objective (para 25). Software whose sole purpose is to archive, collect and transmit data, to name the generic drug or to indicate the contraindications listed in the drug manufacturer's instructions for use does not (para 26).
It does not matter whether the software acts directly in or on the human body (paras 28 and 32). Where it is a medical device, it must bear the CE marking when placed on the market and may then circulate freely in the EU for that function without any additional certification (para 35). For software made up of several modules, only the modules meeting the definition fall within the Directive, and the manufacturer must identify them (paras 36 to 38).
The judgment interprets Directive 93/42/EEC, which Art. 122 of the MDR repeals. Under Art. 2(12) MDR, the intended purpose is determined also by the data supplied by the manufacturer in promotional or sales materials or statements. Caidera checks your assets against CJEU Snitem and flags which wording is risky and why.
CJEU Snitem: key holdings
- Operative part; paras 34 and 39
- Software of which at least one function makes it possible to use patient-specific data to detect, inter alia, contraindications, drug interactions and excessive doses is, in respect of that function, a medical device, even if it does not act directly in or on the human body.
- Paras 22 to 24
- Software must meet a purpose condition and an action condition cumulatively. Use in a medical context is not enough. The manufacturer must have defined a specifically medical intended purpose.
- Para 26
- Software is not a medical device where, although intended for use in a medical context, its sole purpose is to archive, collect and transmit data, to name the generic drug or to indicate the contraindications listed in the drug manufacturer's instructions for use.
- Paras 28 to 32
- The definition does not require the product to act directly in or on the human body. What matters is the purpose of use, not the manner in which the effect materialises.
- Paras 35 to 38
- Software that is a medical device needs the CE marking and may then circulate freely without additional certification. For modular software, only the modules meeting the definition are subject to the Directive, and the manufacturer must identify them.
CJEU Snitem: frequently asked questions
Is data storage software a medical device?
Under para 26 it is not, where its sole purpose in a medical context is to archive, collect and transmit data. The judgment gives patient medical data storage software as an example.
Does software have to act on the body to be a medical device?
No. Under para 28 the definition does not require a product to act directly in or on the human body. Under para 32 the essential point is that the software serves one of the medical purposes in the definition.
What applies to software with several modules?
Only the modules that meet the definition of a medical device fall within the Directive and must bear the CE marking (para 36). The manufacturer must identify which modules are medical devices (para 38).
What role does advertising play for the intended purpose?
In the judgment, the CJEU relies on the intended purpose defined by the manufacturer (para 24). Under Art. 2(12) of the MDR, the intended purpose is the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements and as specified in the clinical evaluation.