Medical Devices Act (FI)Finnish Act on Medical Devices

Medical Devices Act (FI) in brief

National supplement to the MDR and IVDR. Section 10 extends the ban on misleading claims in Art. 7 MDR and IVDR to all marketing and requires the manufacturer, the product name, a statement that it is a CE-marked medical device and, where applicable, the notified body number.

Medical Devices Act (FI) in marketing approval

The Finnish Act on Medical Devices (Laki lääkinnällisistä laitteista, 719/2021) supplements the MDR and IVDR. Section 10 is the key provision for advertising. For claims in labelling, instructions for use, making available, putting into service and advertising, it refers to Art. 7 MDR and Art. 7 IVDR, and it makes clear that their bans on prohibited claims apply to all marketing of the device (Section 10(1)).

On top of that, Section 10 sets national mandatory information to check in every asset: the manufacturer and the product or trade name given at registration (paragraph 2), and a statement that the product is a CE-marked medical device. If the device needs a notified body certificate, the number of that notified body must be stated as well (paragraph 3). A product that is not a medical device must not be claimed to be one in marketing (paragraph 4).

A device intended to be placed on the market as a medical device whose conformity has not yet been demonstrated may be marketed if the marketing makes clear that it is not a CE-marked medical device (paragraph 5). Marketing to consumers is also subject to the Consumer Protection Act (paragraph 6). The authority Fimea may issue rules on procedures in marketing (paragraph 7).

Where marketing breaches Section 10 or Art. 7 MDR or IVDR, Fimea can ban its continuation or renewal and, where safety requires it, order a correction (Section 42). Intentional or grossly negligent breaches are punishable by a fine (Section 57(1) point 3). Caidera checks your assets against Section 10 of the Finnish Medical Devices Act and flags which wording is risky and why.

Medical Devices Act (FI): key provisions

Section 10(1) Medical Devices Act (FI)
Claims in labelling, instructions for use and advertising are governed by Art. 7 MDR and IVDR. The bans set out there apply to all marketing of the device.
Section 10(2) and (3) Medical Devices Act (FI)
Marketing states the manufacturer, the product or trade name given at registration and that the product is a CE-marked medical device. If a notified body certificate is required, also the number of that notified body.
Section 10(4) and (5) Medical Devices Act (FI)
A product must not be marketed as a medical device if it is not one. Devices whose conformity is not yet demonstrated and devices with a derogation under Section 58 may be marketed if the marketing makes clear that they are not CE-marked.
Section 5(1) Medical Devices Act (FI)
Information required for safe use must be in Finnish and Swedish. If the device is intended for patients or other consumers, this also applies to the instructions for use.
Sections 42, 57 Medical Devices Act (FI)
Fimea can ban marketing that breaches Section 10 or Art. 7 MDR or IVDR and, where safety requires it, order a correction. Intentional or grossly negligent breaches of these provisions are punishable by a fine.

Medical Devices Act (FI): advertising examples

  • Not allowedMedical device against tension

    Product is not a medical device

    A product that is not a medical device must not be marketed as one (Section 10(4)).

  • Not allowedComing soon: our new medical device

    Conformity not yet demonstrated, no statement that it is not CE-marked

    Before conformity is demonstrated, marketing is only allowed if it makes clear that the product is not a CE-marked medical device (Section 10(5)).

  • Usually allowedCE-marked medical device

    Ad naming manufacturer, product name and notified body number

    This statement is in fact mandatory (Section 10(3)), as are the manufacturer and product name (Section 10(2)).

  • RiskyEliminates back pain for good

    Under Art. 7 MDR, to which Section 10(1) refers, marketing must not ascribe functions or properties the device does not have.

Simplified examples for orientation, not legal advice. Product, audience and context always decide.

Medical Devices Act (FI): frequently asked questions

What must medical device advertising in Finland state?

Under Section 10(2) and (3): the manufacturer, the product or trade name given at registration and a statement that it is a CE-marked medical device. If a notified body certificate is required, also the number of that notified body.

Can a device be advertised in Finland before CE marking?

Yes, if the device is intended to be placed on the market as a medical device and the marketing makes clear that it is not yet a CE-marked medical device (Section 10(5)). The same applies to devices with a Fimea derogation under Section 58.

Who supervises medical device advertising in Finland?

The authority Fimea (Finnish Medicines Agency). It can ban marketing that breaches Section 10 or Art. 7 MDR or IVDR and, where safety requires it, order a correction (Section 42).