OUG 46/2021Emergency Ordinance No. 46/2021 implementing the Medical Device Regulation
OUG 46/2021 in brief
National advertising rules for medical devices alongside the MDR: public advertising only for devices used without qualified medical staff, mandatory information, banned content and an ANMDMR advertising visa before dissemination. Professional advertising for high-risk devices must be notified in advance.
OUG 46/2021 in marketing approval
Emergency Ordinance No. 46/2021 sets the institutional framework for the MDR in Romania, and its Chapter III regulates medical device advertising. It covers any information by direct contact and any promotion intended to stimulate distribution, sale or use, including visits by representatives, samples and sponsorship of congresses (Art. 17). Labelling and instructions for use are not covered.
All advertising must encourage rational use, must not mislead and must contain correct, up-to-date and verifiable information (Art. 19(1)). Advertising is banned for a device whose way of achieving its intended purpose cannot be supported by clinical evidence (Art. 19(2)). All information must match the instructions for use (Art. 19(3)).
Before approving a consumer asset, check whether the device may be advertised at all: only devices intended for use without qualified medical staff, where a pharmacist's advice is sufficient (Art. 19(5)). Public advertising is banned for devices used only on medical recommendation (Art. 19(4)). The asset needs the name, the intended purpose, information for correct use and an invitation to read the instructions for use (Art. 20(1)). Art. 20(2) bans, among others, guaranteed diagnoses or effects, comparisons with medicines and endorsements by well-known persons. The manufacturer, importer or distributor submits the material to ANMDMR and may disseminate it only after the advertising visa has been issued (Art. 21(1)).
Professional advertising must state that it is intended exclusively for healthcare professionals and give, among other things, the device class and intended purpose (Art. 22). For high-risk devices it must be notified to ANMDMR in advance (Art. 23(1)). Breaches of Art. 19 to 22 are fined 10,000 to 20,000 lei (Art. 28(am), Art. 29(33)). Caidera checks your assets against OUG 46/2021 and flags which wording is risky and why.
OUG 46/2021: key provisions
- Art. 17, 18 OUG 46/2021
- Broad definition of advertising, including representative visits, samples and sponsorship; labelling, instructions for use and replies to specific questions are excluded. ANMDMR is competent for evaluation, notification and approval, except for professional advertising of devices not listed as high-risk.
- Art. 19 OUG 46/2021
- Rational use, no misleading content, correct and verifiable information consistent with the instructions for use. No advertising without clinical evidence for the mode of action. Public advertising only for devices used without qualified medical staff; banned for devices used only on medical recommendation.
- Art. 20 OUG 46/2021
- Public advertising must clearly identify the product as a medical device and include the name, intended purpose, information for use and the invitation to read the instructions for use. Banned are, among others, suggesting a medical consultation is unnecessary, guaranteed results, advertising aimed exclusively at children and natural origin as the reason for safety.
- Art. 21 OUG 46/2021
- Public advertising only after an ANMDMR advertising visa, requested by the manufacturer, importer or distributor with proof of the fee. Assessment within 30 days of a complete dossier, up to 60 days for complex material. If requested documents are not supplied within 30 days, the material is deemed rejected.
- Art. 22, 23 OUG 46/2021
- Professional advertising states that it is exclusively for healthcare professionals and gives the name, manufacturer, intended purpose, characteristics, class, essential information for use and the invitation to read the instructions and label. For high-risk devices it must be notified in advance. ANMDMR can stop advertising and require a corrective statement.
- Art. 28(am), Art. 29(33) OUG 46/2021
- Breaches of Art. 19 to 22 by the manufacturer or the issuer of the promotional material are administrative offences. Fine of 10,000 to 20,000 lei and a request to stop the advertising.
OUG 46/2021: advertising examples
Not allowed
Error-free measurement, guaranteed
Public advertising for a measuring device
Public advertising must not suggest that the result obtained with the device is guaranteed or cannot be subject to error (Art. 20(2)(b)).
Not allowed
Recommended by dermatologists
Public advertising
Reference to recommendations by scientists, healthcare professionals or well-known persons is banned (Art. 20(2)(e)).
Not allowed
Works as well as a painkiller
Public advertising
Advertising must not suggest that the effect is better than or equivalent to treatment with another device or a medicine (Art. 20(2)(b)).
Risky
Clinically proven relief
Only allowed if the mode of action is supported by clinical evidence and the claim matches the instructions for use (Art. 19(2) and (3)).
Simplified examples for orientation, not legal advice. Product, audience and context always decide.
OUG 46/2021: frequently asked questions
Does medical device advertising in Romania need approval?
Public advertising does: it may only be disseminated after ANMDMR has issued the advertising visa (Art. 21(1) OUG 46/2021). The visa under Art. 21 concerns public advertising only; professional advertising for high-risk devices must be notified in advance (Art. 23(1)). ANMDMR reviews professional advertising afterwards and can stop it (Art. 23(2) and (3)).
Which medical devices may be advertised to the public in Romania?
Only for devices that, by their intended purpose, are used without qualified medical staff for diagnosis, recommendation or monitoring of treatment, where a pharmacist's advice is sufficient if needed (Art. 19(5)). Public advertising is banned for special-purpose devices used only on medical recommendation (Art. 19(4)).
What must professional advertising for medical devices contain in Romania?
Under Art. 22: a statement that it is intended exclusively for healthcare professionals, the device name and the manufacturer or its representative, the intended purpose, characteristics and performance, the device class, the information essential for use and an invitation to read the instructions for use and label carefully.